{"id":5723,"date":"2026-11-17T12:41:11","date_gmt":"2026-11-17T11:41:11","guid":{"rendered":"https:\/\/tetravision.be\/?post_type=case&#038;p=5723"},"modified":"2026-11-17T12:41:20","modified_gmt":"2026-11-17T11:41:20","slug":"insulin-pen-ct-scan-dose-accuracy","status":"publish","type":"case","link":"https:\/\/tetravision.be\/en\/case\/insulin-pen-ct-scan-dose-accuracy\/","title":{"rendered":"How much insulin does an insulin pen deliver? We measured it with X-ray CT, without opening the pen"},"content":{"rendered":"\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/insulinepen-ct-scan-cover.png\"><img fetchpriority=\"high\" decoding=\"async\" width=\"1024\" height=\"565\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/insulinepen-ct-scan-cover-1024x565.png\" alt=\"CT scan of a Lantus insulin pen with a 3D model of the cartridge and plunger\" class=\"wp-image-5707\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/insulinepen-ct-scan-cover-1024x565.png 1024w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/insulinepen-ct-scan-cover-300x165.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/insulinepen-ct-scan-cover-768x424.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/insulinepen-ct-scan-cover-650x358.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/insulinepen-ct-scan-cover-600x331.png 600w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/insulinepen-ct-scan-cover.png 1536w\" sizes=\"(max-width: 1024px) 100vw, 1024px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>CT scan of the complete Lantus insulin pen, with the 3D model of the cartridge and plunger.<\/em><\/figcaption><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">An insulin pen is a small mechanical marvel. You dial in a dose, press the button, and a rubber plunger moves just over a millimetre forward in the cartridge. That one millimetre determines how much insulin the patient receives. But the entire mechanism is sealed inside plastic: if you want to see what happens inside, you normally have to break the pen open.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">In this internal study, we used high-resolution X-ray CT to look inside a Lantus SoloStar insulin pen, before and after delivering 10 units. We measured how far the plunger moved, converted that into a volume and compared the result with the standard. The difference from the dialled dose: \u22123.8%, well within the permitted margin. Without disassembling a single part.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Project at a glance<\/h2>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Type<\/strong><\/td><td>Internal study, example case for medical devices<\/td><\/tr><tr><td><strong>Object<\/strong><\/td><td>Lantus SoloStar insulin pen (U-100), complemented by a Humalog insulin pen<\/td><\/tr><tr><td><strong>Research question<\/strong><\/td><td>Can X-ray CT measure the mechanical dose delivery through the displacement of the plunger?<\/td><\/tr><tr><td><strong>Reference<\/strong><\/td><td>ISO 11608-1:2022, dose accuracy requirements for pen injectors<\/td><\/tr><tr><td><strong>Technique<\/strong><\/td><td><a href=\"https:\/\/tetravision.be\/en\/x-ray-ct-scanning\/\">X-ray CT scanning<\/a> (Comet Yxlon FF35, helical CT)<\/td><\/tr><tr><td><strong>Voxel size<\/strong><\/td><td>19.5 \u00b5m<\/td><\/tr><tr><td><strong>Analysis<\/strong><\/td><td>ZEISS INSPECT X-Ray<\/td><\/tr><tr><td><strong>Result<\/strong><\/td><td>Plunger displacement of 1.339 mm = approx. 9.62 units at a dialled dose of 10 units (\u22123.8%), within the ISO limit of \u00b11 unit<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\">The challenge: checking a dose you cannot see<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">For pen injectors, ISO 11608-1:2022 specifies how accurate the delivered dose must be. The permitted deviation depends on the dialled dose and on the dose resolution of the pen. For a pen that can be set in 1-unit steps, a maximum deviation of \u00b11 unit applies up to 20 units. For higher doses, it is \u00b15% of the dialled dose.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/iso-11608-1-dosisnauwkeurigheid-grafiek.png\"><img decoding=\"async\" width=\"807\" height=\"402\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/iso-11608-1-dosisnauwkeurigheid-grafiek.png\" alt=\"Chart of the dose accuracy specification limits according to ISO 11608-1\" class=\"wp-image-5708\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/iso-11608-1-dosisnauwkeurigheid-grafiek.png 807w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/iso-11608-1-dosisnauwkeurigheid-grafiek-300x149.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/iso-11608-1-dosisnauwkeurigheid-grafiek-768x383.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/iso-11608-1-dosisnauwkeurigheid-grafiek-650x324.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/iso-11608-1-dosisnauwkeurigheid-grafiek-600x299.png 600w\" sizes=\"(max-width: 807px) 100vw, 807px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>Dose accuracy specification limits according to ISO 11608-1. Source: <a href=\"https:\/\/pmc.ncbi.nlm.nih.gov\/articles\/PMC5851229\/\" target=\"_blank\" rel=\"noopener\">PMC5851229 (NCBI)<\/a>.<\/em><\/figcaption><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">For this measurement, we dialled 10 units, or 0.10 ml of insulin (U-100 means 100 units per ml). The actual delivered dose may therefore lie between 9 and 11 units, or between 0.09 and 0.11 ml.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">The problem: the parts that determine that dose sit deep inside the pen. Taking the pen apart disturbs exactly what you want to measure. X-ray CT looks straight through it. The movement of the plunger becomes visible and measurable, while the pen stays intact and can still be used.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">The research question<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Can X-ray CT quantify the mechanical dose delivery of an insulin pen, based on the displacement of the plunger and the geometry of the cartridge? We split that into two questions:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li><strong>Plunger displacement.<\/strong> Can we accurately determine the position of the rubber plunger before and after delivering 10 units?<\/li>\n\n\n\n<li><strong>Volume determination.<\/strong> Can we convert that displacement, together with the inner diameter of the cartridge, into the displaced volume of insulin?<\/li>\n<\/ul>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-mechanisme-voor-na-injectie.png\"><img decoding=\"async\" width=\"1024\" height=\"332\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-mechanisme-voor-na-injectie-1024x332.png\" alt=\"Diagram of the mechanism of a Lantus insulin pen before and after injection\" class=\"wp-image-5709\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-mechanisme-voor-na-injectie-1024x332.png 1024w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-mechanisme-voor-na-injectie-300x97.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-mechanisme-voor-na-injectie-768x249.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-mechanisme-voor-na-injectie-1536x498.png 1536w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-mechanisme-voor-na-injectie-650x211.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-mechanisme-voor-na-injectie-600x194.png 600w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-mechanisme-voor-na-injectie.png 1849w\" sizes=\"(max-width: 1024px) 100vw, 1024px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>The internal mechanism of the Lantus insulin pen. After the injection, the plunger has moved forward (z\u2081 \u2192 z\u2082). In magenta, the inner diameter in which the plunger slides.<\/em><\/figcaption><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\">Our approach<\/h2>\n\n\n\n<ol class=\"wp-block-list\">\n<li><strong>Two scans of the same pen.<\/strong> We scanned the pen before the injection, then delivered 10 units and scanned the pen again. Scan, reconstruction and measurement settings were identical for both scans, otherwise you are comparing apples with oranges.<\/li>\n\n\n\n<li><strong>Helical CT at high magnification.<\/strong> We focused the measurement on the cartridge and the plunger. With a helical acquisition on our Comet Yxlon FF35, we capture a long section of the cartridge while the pen sits close to the X-ray source for high magnification. This gave us a voxel size of 19.5 \u00b5m.<\/li>\n<\/ol>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-scan-voor-na-injectie.png\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"610\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-scan-voor-na-injectie-1024x610.png\" alt=\"CT scan of a Lantus insulin pen before and after injection\" class=\"wp-image-5710\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-scan-voor-na-injectie-1024x610.png 1024w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-scan-voor-na-injectie-300x179.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-scan-voor-na-injectie-768x458.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-scan-voor-na-injectie-1536x915.png 1536w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-scan-voor-na-injectie-650x387.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-scan-voor-na-injectie-600x358.png 600w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-scan-voor-na-injectie.png 1965w\" sizes=\"(max-width: 1024px) 100vw, 1024px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>CT image of the complete Lantus insulin pen, (a) before and (b) after the injection, with the detail zones around the plunger.<\/em><\/figcaption><\/figure>\n\n\n\n<ol start=\"3\" class=\"wp-block-list\">\n<li><strong>Choosing a fixed reference.<\/strong> In ZEISS INSPECT X-Ray, we built a 3D model of the zone around the cartridge and plunger. That zone contains a part that does not move, the plunger holder, and a part that does, the plunger itself. The plunger holder serves as a fixed reference to align both scans.<\/li>\n<\/ol>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-plunjerhouder-plunjer.png\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"583\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-plunjerhouder-plunjer-1024x583.png\" alt=\"CT image of the plunger holder and plunger in an insulin pen\" class=\"wp-image-5711\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-plunjerhouder-plunjer-1024x583.png 1024w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-plunjerhouder-plunjer-300x171.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-plunjerhouder-plunjer-768x437.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-plunjerhouder-plunjer-1536x874.png 1536w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-plunjerhouder-plunjer-650x370.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-plunjerhouder-plunjer-600x341.png 600w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-ct-plunjerhouder-plunjer.png 1912w\" sizes=\"(max-width: 1024px) 100vw, 1024px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>The zone we analysed, with the fixed plunger holder and the moving plunger.<\/em><\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-cartridge-plunjer.png\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"676\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-cartridge-plunjer-1024x676.png\" alt=\"3D model of the cartridge and plunger of an insulin pen\" class=\"wp-image-5712\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-cartridge-plunjer-1024x676.png 1024w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-cartridge-plunjer-300x198.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-cartridge-plunjer-768x507.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-cartridge-plunjer-650x429.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-cartridge-plunjer-600x396.png 600w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-cartridge-plunjer.png 1364w\" sizes=\"(max-width: 1024px) 100vw, 1024px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>3D model of the cartridge and plunger.<\/em><\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-doorsnede.png\"><img loading=\"lazy\" decoding=\"async\" width=\"968\" height=\"549\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-doorsnede.png\" alt=\"Cross-section of the 3D model of the cartridge and plunger\" class=\"wp-image-5713\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-doorsnede.png 968w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-doorsnede-300x170.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-doorsnede-768x436.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-doorsnede-650x369.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-3d-model-doorsnede-600x340.png 600w\" sizes=\"(max-width: 968px) 100vw, 968px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>Cross-section of the 3D model.<\/em><\/figcaption><\/figure>\n\n\n\n<ol start=\"4\" class=\"wp-block-list\">\n<li><strong>Measuring the inner diameter.<\/strong> On a cross-section of the 3D model, we measured the inner diameter of the cartridge in which the plunger slides: 9.563 mm.<\/li>\n<\/ol>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-cartridge-3d-model.png\"><img loading=\"lazy\" decoding=\"async\" width=\"852\" height=\"1024\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-cartridge-3d-model-852x1024.png\" alt=\"Measurement of the inner diameter of the insulin cartridge on the 3D model\" class=\"wp-image-5714\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-cartridge-3d-model-852x1024.png 852w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-cartridge-3d-model-250x300.png 250w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-cartridge-3d-model-768x923.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-cartridge-3d-model-541x650.png 541w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-cartridge-3d-model-600x721.png 600w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-cartridge-3d-model.png 862w\" sizes=\"(max-width: 852px) 100vw, 852px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>The inner diameter of the cartridge (9.563 mm), measured on a cross-section of the 3D model.<\/em><\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-ct-slice.png\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"129\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-ct-slice-1024x129.png\" alt=\"Inner diameter of the cartridge marked on a CT slice of the insulin pen\" class=\"wp-image-5715\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-ct-slice-1024x129.png 1024w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-ct-slice-300x38.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-ct-slice-768x97.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-ct-slice-1536x194.png 1536w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-ct-slice-650x82.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-ct-slice-600x76.png 600w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-binnendiameter-ct-slice.png 1963w\" sizes=\"(max-width: 1024px) 100vw, 1024px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>The same diameter, marked in magenta on the CT slice.<\/em><\/figcaption><\/figure>\n\n\n\n<ol start=\"5\" class=\"wp-block-list\">\n<li><strong>Measuring the displacement.<\/strong> We measured the distance between the plunger holder and the plunger in both states. The difference is the displacement of the plunger: \u0394z = z\u2082 \u2212 z\u2081.<\/li>\n\n\n\n<li><strong>Converting to a dose.<\/strong> Displacement and diameter together give the displaced volume: \u0394V \u2248 \u03c0 \u00b7 (d\/2)\u00b2 \u00b7 \u0394z. We convert that to ml and then to units of insulin. In short: dialled dose \u2192 plunger displacement (mm) \u2192 volume (mm\u00b3) \u2192 calculated dose (units).<\/li>\n<\/ol>\n\n\n\n<p class=\"wp-block-paragraph\">Read more about how we choose a scanning strategy for a part like this in <a href=\"https:\/\/tetravision.be\/en\/which-ct-scanning-strategy-is-right-for-your-part\/\">Which CT scanning strategy is right for your part?<\/a><\/p>\n\n\n\n<h2 class=\"wp-block-heading\">The result<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Before the injection, the distance between the plunger holder and the plunger was 2.692 mm, afterwards 4.031 mm. The plunger therefore moved 1.339 mm. With an inner diameter of 9.563 mm, that gives a displaced volume of approximately 0.0962 ml, or 9.62 units.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-afstand-plunjer-voor-na-injectie.png\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"430\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-afstand-plunjer-voor-na-injectie-1024x430.png\" alt=\"Measurement of the distance between plunger holder and plunger before and after injection\" class=\"wp-image-5716\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-afstand-plunjer-voor-na-injectie-1024x430.png 1024w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-afstand-plunjer-voor-na-injectie-300x126.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-afstand-plunjer-voor-na-injectie-768x322.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-afstand-plunjer-voor-na-injectie-1536x644.png 1536w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-afstand-plunjer-voor-na-injectie-650x273.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-afstand-plunjer-voor-na-injectie-600x252.png 600w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-afstand-plunjer-voor-na-injectie.png 1845w\" sizes=\"(max-width: 1024px) 100vw, 1024px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>Distance between the plunger holder and the plunger, (a) before and (b) after the injection.<\/em><\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-plunjerverplaatsing.png\"><img loading=\"lazy\" decoding=\"async\" width=\"866\" height=\"812\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-plunjerverplaatsing.png\" alt=\"Overlay of the plunger position before and after injection\" class=\"wp-image-5717\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-plunjerverplaatsing.png 866w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-plunjerverplaatsing-300x281.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-plunjerverplaatsing-768x720.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-plunjerverplaatsing-650x609.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/lantus-insulinepen-plunjerverplaatsing-600x563.png 600w\" sizes=\"(max-width: 866px) 100vw, 866px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>Both cross-sections aligned on the fixed plunger holder. The difference between blue and red is the movement of the plunger.<\/em><\/figcaption><\/figure>\n\n\n\n<figure class=\"wp-block-table\"><table class=\"has-fixed-layout\"><tbody><tr><td><strong>Dialled dose<\/strong><\/td><td>10 units (0.100 ml)<\/td><\/tr><tr><td><strong>Distance before injection<\/strong><\/td><td>2.692 mm<\/td><\/tr><tr><td><strong>Distance after injection<\/strong><\/td><td>4.031 mm<\/td><\/tr><tr><td><strong>Plunger displacement \u0394z<\/strong><\/td><td>1.339 mm<\/td><\/tr><tr><td><strong>Inner diameter of cartridge<\/strong><\/td><td>9.563 mm<\/td><\/tr><tr><td><strong>Calculated volume<\/strong><\/td><td>approx. 0.0962 ml<\/td><\/tr><tr><td><strong>Calculated dose<\/strong><\/td><td>approx. 9.62 units<\/td><\/tr><tr><td><strong>Deviation<\/strong><\/td><td>approx. \u22123.8%<\/td><\/tr><tr><td><strong>Permitted under ISO 11608-1<\/strong><\/td><td>9 to 11 units<\/td><\/tr><\/tbody><\/table><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">The calculated dose falls within the tolerances of ISO 11608-1:2022. More important for us: CT proves to be a suitable, non-destructive method to visualise how an insulin pen works and to measure the displacement of the plunger quantitatively.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\"><strong>One caveat.<\/strong> This study covers one pen and one dose. A volume calculated from CT is also not the same as a direct measurement of the liquid leaving the needle. It is a strong, independent check of the mechanism, not a replacement for the gravimetric test in the standard.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">What else CT reveals<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Besides the Lantus pen, we also scanned a Humalog insulin pen. Internally, it is built in a similar way: a large number of plastic parts around a cartridge of insulin.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-doorsnede.png\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"128\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-doorsnede-1024x128.png\" alt=\"CT cross-section of a Humalog insulin pen\" class=\"wp-image-5718\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-doorsnede-1024x128.png 1024w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-doorsnede-300x38.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-doorsnede-768x96.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-doorsnede-650x81.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-doorsnede-600x75.png 600w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-doorsnede.png 1429w\" sizes=\"(max-width: 1024px) 100vw, 1024px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>CT cross-section of the complete Humalog insulin pen.<\/em><\/figcaption><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">At the back sits a metal spring. Metal causes typical artefacts in a CT scan, so the surrounding plastic parts appear less sharp locally. Even there, the scan still contains a lot of useful information.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-veer-metaalartefacten.png\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"568\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-veer-metaalartefacten-1024x568.png\" alt=\"CT detail of the metal spring in an insulin pen with metal artefacts\" class=\"wp-image-5719\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-veer-metaalartefacten-1024x568.png 1024w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-veer-metaalartefacten-300x167.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-veer-metaalartefacten-768x426.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-veer-metaalartefacten-650x361.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-veer-metaalartefacten-600x333.png 600w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-veer-metaalartefacten.png 1405w\" sizes=\"(max-width: 1024px) 100vw, 1024px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>The back of the pen with the metal spring and the resulting metal artefacts.<\/em><\/figcaption><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">In the middle section, you can see how the plastic parts interlock and work together.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-middengedeelte.png\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"594\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-middengedeelte-1024x594.png\" alt=\"CT detail of the plastic parts in the middle of an insulin pen\" class=\"wp-image-5720\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-middengedeelte-1024x594.png 1024w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-middengedeelte-300x174.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-middengedeelte-768x446.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-middengedeelte-650x377.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-middengedeelte-600x348.png 600w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-middengedeelte.png 1285w\" sizes=\"(max-width: 1024px) 100vw, 1024px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>The middle section with the interacting plastic parts.<\/em><\/figcaption><\/figure>\n\n\n\n<p class=\"wp-block-paragraph\">At the front, the details of the cartridge and needle come into view, and even a few small air bubbles in the insulin.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><a href=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-naald-luchtbelletjes.png\"><img loading=\"lazy\" decoding=\"async\" width=\"1024\" height=\"494\" src=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-naald-luchtbelletjes-1024x494.png\" alt=\"CT detail of the needle and air bubbles in an insulin cartridge\" class=\"wp-image-5721\" srcset=\"https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-naald-luchtbelletjes-1024x494.png 1024w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-naald-luchtbelletjes-300x145.png 300w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-naald-luchtbelletjes-768x370.png 768w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-naald-luchtbelletjes-650x313.png 650w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-naald-luchtbelletjes-600x289.png 600w, https:\/\/tetravision.be\/wp-content\/uploads\/2026\/11\/humalog-insulinepen-ct-naald-luchtbelletjes.png 1431w\" sizes=\"(max-width: 1024px) 100vw, 1024px\" \/><\/a><figcaption class=\"wp-element-caption\"><em>The front with cannula, needle and air bubbles in the insulin.<\/em><\/figcaption><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\">Why this matters for medical devices<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">An insulin pen is just one example of a closed dosing system. The same approach works for other medical devices where small mechanical movements make a big difference. We use industrial CT there for, among other things:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li>non-destructive inspection of complete assemblies<\/li>\n\n\n\n<li>dimensional metrology of internal parts<\/li>\n\n\n\n<li>functional analysis, such as the position of parts before and after use<\/li>\n\n\n\n<li>volume determination in cartridges, reservoirs and channels<\/li>\n\n\n\n<li><a href=\"https:\/\/tetravision.be\/en\/quality-inspection\/\">quality inspection<\/a> and comparing production batches<\/li>\n<\/ul>\n\n\n\n<p class=\"wp-block-paragraph\">In an earlier case, we used the same approach to reveal a <a href=\"https:\/\/tetravision.be\/en\/case\/detecting-leaks-without-disassembly\/\">leaking valve system in a bottle<\/a>, again without disassembling anything.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Working on a medical device?<\/h2>\n\n\n\n<p class=\"wp-block-paragraph\">Do you develop or manufacture injection systems, inhalers, dosing pumps or other medical devices, and want to know what happens inside? We are happy to help you find the right scan and analysis. Contact us at <a href=\"mailto:info@tetravision.be\">info@tetravision.be<\/a> or <a href=\"tel:+3216910420\">+32 16 91 04 20<\/a>. We always respond within 24 hours.<\/p>\n\n\n\n<p class=\"wp-block-paragraph\">Want to know more about our CT capabilities? Visit our <a href=\"https:\/\/tetravision.be\/en\/x-ray-ct-scanning\/\">X-ray CT scanning page<\/a>.<\/p>\n\n","protected":false},"excerpt":{"rendered":"<p>An insulin pen is a small mechanical marvel. You dial in a dose, press the button, and a rubber plunger moves just over a millimetre forward in the cartridge. That one millimetre determines how much insulin the patient receives. But the entire mechanism is sealed inside plastic: if you want to see what happens inside, [&hellip;]<\/p>\n","protected":false},"featured_media":5707,"template":"","meta":{"_acf_changed":false},"case-categorie":[20],"class_list":["post-5723","case","type-case","status-publish","has-post-thumbnail","hentry","case-categorie-xray-ct-scanning-en"],"acf":[],"_links":{"self":[{"href":"https:\/\/tetravision.be\/en\/wp-json\/wp\/v2\/case\/5723","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/tetravision.be\/en\/wp-json\/wp\/v2\/case"}],"about":[{"href":"https:\/\/tetravision.be\/en\/wp-json\/wp\/v2\/types\/case"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/tetravision.be\/en\/wp-json\/wp\/v2\/media\/5707"}],"wp:attachment":[{"href":"https:\/\/tetravision.be\/en\/wp-json\/wp\/v2\/media?parent=5723"}],"wp:term":[{"taxonomy":"case-categorie","embeddable":true,"href":"https:\/\/tetravision.be\/en\/wp-json\/wp\/v2\/case-categorie?post=5723"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}